Group
Vertical
Syntax
CH<x>:SCAle <NR3>
CH<x>:SCAle?
Related Commands
CH<x>:OFFSet on page 63, CH<x>:POSition on page 64,
REF<x>:VERTical:SCAle? on page 203, MATH:VERtical:SCAle on page 169
Arguments
<NR3> is the scale, in units-per-division. The value entered here is truncated to
three significant digits.
Examples
CH1:SCALE 100E-3 sets the channel 1 gain to 100 mV/div.
CH2:SCALE? might return :CH2:SCALE 1.0000, indicating that the current V/
div setting of channel 2 is 1 V/div.
CH<x>:VOLts
Sets or queries the vertical sensitivity of the specified channel. The value of <x>
can vary from 1 through 4 depending on the instrument model.
This command is identical to the CH<x>:SCAle command and is included for
compatibility purposes. Only CH<x>:SCAle is returned in response to a CH<x>?
query.
Group
Vertical
Syntax
CH<x>:VOLts <NR3>
CH<x>:VOLts?
C commands
TBS2000 Series Programmer 73