EasyManua.ls Logo

Tektronix TBS2000 Series User Manual

Tektronix TBS2000 Series
328 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #204 background imageLoading...
Page #204 background image
Group
Measurement
Syntax
MEASUrement:MEAS<x>:STATE {OFF|ON|<NR1>}
MEASUrement:MEAS<x>:STATE?
Related commands
MEASUrement:MEAS<x>:SOUrce1 on page 183, MEASUrement:MEAS<x>:
TYPe on page 186
Arguments
OFF disables calculation and display of the specified measurement slot.
ON enables calculation and display of the specified measurement slot.
<NR1> = 0 disables calculation and display of the specified measurement slot;
any other value enables calculation and display of the specified measurement
slot.
Examples
MEASUrement:MEAS2:STATE ON computes and displays the measurement
defined as measurement 2.
MEASUrement:MEAS1:STATE? might
return :MEASUREMENT:MEAS1:STATE 0 indicating that measurement
defined for measurement slot 1 is disabled.
MEASUrement:MEAS<x>:TYPe
Sets or queries the on-screen periodic instrument measurement type for the
measurement specified by <x>. Where <x> identifies the measurement, 1 through
6 depending on instrument model.
This is equivalent to selecting the measurement type in the MEASURE menu.
Setting the type to anything other than NONE displays the MEASURE menu on
the screen.
NOTE. You should use the MEASUrement:IMMed command with programming to
take measurements, as this is preferred to the
MEASUrement:MEAS<x>command.
M commands
186 TBS2000 Series Programmer

Table of Contents

Other manuals for Tektronix TBS2000 Series

Question and Answer IconNeed help?

Do you have a question about the Tektronix TBS2000 Series and is the answer not in the manual?

Tektronix TBS2000 Series Specifications

General IconGeneral
Bandwidth70 MHz, 100 MHz, 200 MHz
Channels2 or 4
Time Base Range5 ns/div to 50 s/div
Probe InterfaceTekVPI
Operating Temperature0 °C to +50 °C
Vertical Resolution8 bits
Sample Rate1 GS/s
Memory Depth20 kpts
Record Length20 kpts
Display9-inch WVGA display
Input Impedance1 MΩ ±1%, 13 pF ±2 pF
Vertical Sensitivity2 mV/div to 5 V/div
Trigger TypesEdge, Pulse Width, Video
ConnectivityUSB, LAN
Math FunctionsAddition, Subtraction, Multiplication, Division, FFT

Summary

Getting Started with TBS2000 Series Programmer

Remote Operation Overview

Provides information on remotely controlling the instrument via communication protocols and commands.

Understanding Command Syntax

Command and Query Structure

Explains the formation of set and query commands, including headers, mnemonics, and arguments.

BNF Notation

Details the Backus-Naur Form used for describing command syntax and structure.

Command Entry Rules

Outlines general rules for entering commands, including case sensitivity and whitespace.

Concatenating Commands

Explains how to combine multiple commands and queries using semicolons for sequential execution.

Organizing Commands into Groups

Acquisition Command Group

Manages waveform acquisition modes, settings, and averaging parameters.

Ethernet Command Group

Controls the instrument's Ethernet interface settings and operations.

Horizontal Command Group

Manages horizontal axis settings, time base, and record length.

Measurement Command Group

Configures and retrieves automated measurement results.

Trigger Command Group

Defines and manages instrument trigger conditions and settings.

Understanding Status and Events

Registers Overview

Explains the instrument's status and enable registers for event reporting.

Event Handling Sequence

Details the process of how the status and event handling system manages instrument events.

Related product manuals