D commands
This section lists commands and queries that begin with the letter D.
DATa
Sets or queries the format and location of the waveform data that is transferred
with the CURVe command.
Group
Waveform
Syntax
DATa {INIT|SNAp}
DATa?
Related Commands
CURVe on page 87, DATa:STARt on page 92, DATa:STOP on page 93,
WFMInpre:NR_Pt? on page 251, WFMOutpre:NR_Pt? on page 261
Arguments
INIT reinitializes the waveform data settings to their factory defaults except for
DATa:STOP, which is set to the current acquisition record length.
SNAp sets DATa:STARt and DATa:STOP to match the current waveform cursor
positions.
Examples
DATaINIT initializes the waveform data settings to their factory defaults.
DATa? might return :DATA:DESTINATION REF1:ENCDG
RIBINARY;SOURCE CH1;START 1;STOP 500;WIDTH 1.