Group
Measurement
Syntax
MEASUrement:IMMed:VALue?
Related Commands
MEASUrement:IMMed:TYPe on page 177, MEASUrement:IMMed:SOUrce1 on
page 175, *ESR? on page 107, EVENT? on page 115
Returns
<NR3>
Examples
MEASUrement:IMMed:VALue? might return 28.75E6 if you are measuring the
frequency of a 28.75 MHz signal.
MEASUrement:IMMed:VALue? might return 9.9000E+37. If the measurement
has an error or warning associated with it, then an item is added to the error
queue. The error can be checked for with the *ESR? and ALLEv? commands..
MEASUrement:MEAS<x>?
Returns all measurement parameters for the displayed instrument periodic
measurement specified by <x>. Where <x> identifies the measurement, 1 through
6 depending on instrument model. Query only.
Group
Measurement
Syntax
MEASUrement:MEAS<x>?
Returns
Settings for the specified measurement source.