EasyManuals Logo

Tektronix TBS2000 Series User Manual

Tektronix TBS2000 Series
328 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #110 background imageLoading...
Page #110 background image
DATa:STARt
Sets or queries the starting data point for incoming or outgoing waveform
transfer. This command lets you transfer partial waveforms to and from the
instrument.
Group
Waveform
Syntax
DATa:STARt <NR1>
DATa:STARt?
Related Commands
CURVe on page 87, DATa on page 89, DATa:STOP on page 93, WFMInpre:
NR_Pt? on page 251, WFMOutpre:NR_Pt? on page 261
Arguments
<NR1> is the first data point that will be transferred, which ranges from 1 to the
record length. Data will be transferred from <NR1> to DATa:STOP or the record
length, whichever is less. If <NR1> is greater than the record length, the last data
point in the record is transferred. DATa:STARt and DATa:STOP are order
independent. When DATa:STOP is greater than DATa:STARt, the values will be
swapped internally for the CURVE? query.
Examples
DATa:STARt10 specifies that the waveform transfer will begin with data point
10.
DATa:STARt? might return :DATA:START 214 indicating that data point 214 is
the first waveform data point that will be transferred.
D commands
92 TBS2000 Series Programmer

Table of Contents

Other manuals for Tektronix TBS2000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix TBS2000 Series and is the answer not in the manual?

Tektronix TBS2000 Series Specifications

General IconGeneral
BrandTektronix
ModelTBS2000 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals