TRIGger:A:RUNT:POLarity
Sets or queries the polarity for the runt trigger.
Group
Trigger
Syntax
TRIGger:A:RUNT:POLarity {NEGative|POSitive}
TRIGger:A:RUNT:POLarity?
Arguments
POSitive indicates that the rising edge crosses the low threshold and the falling
edge recrosses the low threshold without either edge ever crossing the high
threshold.
NEGative indicates that the falling edge crosses the high threshold and the rising
edge recrosses the high threshold without either edge ever crossing the low
threshold.
Examples
TRIGGER:A:RUNT:POLARITY NEGATIVE specifies that the polarity of the A
pulse runt trigger is negative.
TRIGGER:A:RUNT:POLARITY? might
return :TRIGGER:A:RUNT:POLARITY POSITIVE indicating that the polarity
of the A pulse runt trigger is positive.