WFMInpre:YOFf
Sets or queries the vertical position of the incoming waveform in digitizing
levels. Variations in this number are analogous to changing the vertical position
of the waveform.
Group
Waveform
Syntax
WFMInpre:YOFf <NR3>
WFMInpre:YOFf?
Arguments
<NR3> is a value expressed in digitizing levels.
Examples
WFMINPRE:YOFF 50 specifies that the zero reference point for the incoming
waveform is 50 digitizing levels (2 divisions, for 8-bit data) above the center of
the data range.
WFMINPRE:YOFF? might return :WFMINPRE:YOFF 25 indicating the vertical
position of the incoming waveform in digitizing levels.
WFMInpre:YUNit
Sets or returns the vertical units of the incoming waveform.
Group
Waveform
Syntax
WFMInpre:YUNit <Qstring>
WFMInpre:YUNit?
W commands