CALibrate:FACtory:STATus?
Returns the factory calibration status value saved in nonvolatile memory. Query
only.
Group
Calibration and Diagnostic
Syntax
CALibrate:FACtory:STATus?
Examples
CALIBRATE:FACTORY:STATUS? might
return :CALIBRATE:FACTORY:STATUS PASS indicating that factory
calibration passed.
CALibrate:FACtory:STEPSTIMulus?
Returns the input signal of current factory calibration step. Query only.
Group
Calibration and Diagnostic
Syntax
CALibrate:FACtory:STEPSTIMulus?
Returns
A string that includes the input voltage, input frequency, input impendence, step
number, and step name.
Examples
CALibrate:FACtory:STEPSTIMulus? might return
"0.100000,1000000.000000,1,1e+06,2,L02" indicating:
input voltage: 0.1 V
input frequency : 1000000 Hz
input impendence: 1M
step number: 2
step name: “L02”