EasyManuals Logo

Tektronix TBS2000 Series User Manual

Tektronix TBS2000 Series
328 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #215 background imageLoading...
Page #215 background image
R commands
This section lists commands and queries that begin with the letter R.
*RCL
Restores the state of the instrument from a copy of its settings stored in memory.
(The settings are stored using the *SAV command.) This command is equivalent
to RECAll:SETUp, and performs the same function as the Recall Saved Setup
item in the front-panel Save/Recall Setup menu. Command only, no query form.
Group
Save and Recall
Syntax
*RCL <NR1>
Related Commands
FACtory on page 119, *LRN? on page 164, RECAll:SETUp on page 198, *RST
on page 204, *SAV on page 205, SAVe:SETUp on page 209
Arguments
<NR1> is an integer value in the range from 1 to 10, and specifies a setup storage
location.
Examples
*RCL3 restores the instrument from a copy of the settings stored in memory
location 3.
TBS2000 Series Programmer 197