F commands
This section lists commands and queries that begin with the letter F.
FACtory
Resets the instrument to its factory default settings. Refer to Appendix B: Factory
Setup for a list of the factory default settings. No query.
This command does the following: •
â– 
Clears the Event Status Enable Register
â– 
Clears the Service Request Enable Register
â– 
Sets the Device Event Status Enable Register to 255
â– 
Purges all defined aliases
â– 
Enables all Command Headers
â– 
Sets the macro defined by *DDT to a "zero-length field
â– 
Clears the pending operation flag and associated operations
This command does not reset the following:
â– 
Communication settings
â– 
Sate of the VXI-11 (Ethernet IEEE Std 488.2) interface
â– 
Calibration data that affects device specifications
â– 
Protected user data
â– 
Stored settings
â– 
Power On Status Clear Flag
â– 
instrument password
Group
Save and Recall
Syntax
FACtory
TBS2000 Series Programmer 119

Table of Contents

Other manuals for Tektronix TBS2000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix TBS2000 Series and is the answer not in the manual?

Tektronix TBS2000 Series Specifications

General IconGeneral
BrandTektronix
ModelTBS2000 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals