HORizontal:PREViewstate
Returns a boolean value to indicate whether the acquisition system is in the
preview state. Query only.
Group
Horizontal
Syntax
HORizontal:PREViewstate?
Returns
<NR1> = 1 if the acquisition system is in the preview state.
<NR1> = 0 if the acquisition system is not in the preview state.
Examples
HORizontal:PREViewstate? might return :HORizontal:PREViewstate 1
indicating the acquisition system is in the preview state.
HORizontal:RECOrdlength
Sets the horizontal record length of acquired waveforms. The query form of this
command returns the current horizontal record length.
Group
Horizontal
Syntax
HORizontal:RECOrdlength <NR1>
HORizontal:RECOrdlength?
Arguments
<NR1> represents the supported values for horizontal record lengths, which are:
2000, 20000, 200000, 2000000, 20000000.
Examples
HORIZONTAL:RECORDLENGTH 2000 specifies that 2000 data points will be
acquired for each record.
HORIZONTAL:RECORDLENGTH? might
return :HORIZONTAL:RECOrdlength 2000 indicating that the horizontal record
length is equal to 2000 data points.
H commands
TBS2000 Series Programmer 157

Table of Contents

Other manuals for Tektronix TBS2000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix TBS2000 Series and is the answer not in the manual?

Tektronix TBS2000 Series Specifications

General IconGeneral
BrandTektronix
ModelTBS2000 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals