CALibrate:INTERNAL
This command starts a signal path compensation. Command only, no query form.
NOTE. The self-calibration can take several minutes to complete. During this
time, the instrument does not execute any commands.
Disconnect all signals from the instrument before performing an internal self-
calibration.
Group
Calibration and Diagnostic
Syntax
CALibrate:INTERNAL
Examples
CALibrate:INTERNAL starts a signal path compensation cycle.
CALibrate:INTERNal:STARt
Starts the internal signal path calibration (SPC) of the instrument. You can use
the CALibrate:INTERNal:STATus? query to return the current status of the
internal signal path calibration of the instrument. No query form.
Group
Calibration and Diagnostic
Syntax
CALibrate:INTERNal:STARt
Related commands
CALibrate:RESults:SPC? on page 59
Examples
CALIBRATE:INTERNAL:START initiates the internal signal path
compensation of the instrument.
C commands
56 TBS2000 Series Programmer

Table of Contents

Other manuals for Tektronix TBS2000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix TBS2000 Series and is the answer not in the manual?

Tektronix TBS2000 Series Specifications

General IconGeneral
BrandTektronix
ModelTBS2000 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals