RECAll:SETUp
Restores the factory-default instrument settings, user-saved settings from internal
nonvolatile memory, or user-saved settings from a file on a USB flash drive.
Using the FACTORY argument is equivalent to pushing the DEFAULT SETUP
front-panel button. Command only, no query form.
Group
Save and Recall
Syntax
RECAll:SETUp {FACtory|<NR1>|<file path>}
Related Commands
FACtory on page 119, *RCL on page 197, *RST on page 204, *SAV on
page 205, SAVe:SETUp on page 209, FILESystem:CWD on page 126
Arguments
FACtory selects the factory setup.
<NR1> is a value in the range from 1 to 10, and specifies a setup memory storage
location.
<file path> is a quoted string that defines the path and name of the .SET setup file
to recall from a USB drive. Input the file path using the form <drive>:/<dir>/
<filename>.<extension> and one or more <dir>s are optional. If you do not
specify them, the instrument will read the file from the default directory (see
FILESystem:CWD). <filename> stands for a filename; the use of wildcard
characters in filenames is not supported. Filename extensions are not required,
but highly recommended.
Examples
RECAll:SETUp FACTORY recalls the instrument setup to its factory defaults.
RECALL:SETUP 2 recalls the instrument setup from setup storage location 2.
RECALL:SETUP "TEK00000.SET" recalls the setup from the file
TEK00000.SET in the current directory (such as “usb0/”).
R commands
198 TBS2000 Series Programmer