Arguments
<NR3> is the vertical sensitivity, in volts.
Examples
CH1:VOLts 1.0 sets channel 1 to 1 Volt per division.
CH1:VOLts? Might return CH1:VOLts 1.0 indication that the ch1 volts per
division is 1 Volt per division.
CH<x>:YUNit
Sets or queries the units of the specified channel. The value of <x> can vary from
1 through 4 depending on instrument model number.
String arguments are case insensitive and any unsupported units will generate an
error. Supported units are:
%, /Hz, A, A/A, A/V, A/W, A/dB, A/s, AA, AW, AdB, As, B, Hz, IRE, S/s, V,
V/A, V/V, V/W, V/dB, V/s, VV, VW, VdB, Volts, Vs, W, W/A, W/V, W/W, W/
dB, W/s,WA, WV,WW, WdB, Ws, dB, dB/A, dB/V, dB/W, dB/dB, dBA, dBV,
dBW, dBdB, day, degrees, div, hr, min, ohms, percent, s.
Group
Vertical
Syntax
CH<x>:YUNit <QString>
CH<x>:YUNit?
Arguments
<QString> is a string of text surrounded by quotes, specifying the supported
units. This command is case insensitive.
Examples
CH2:YUNit "V" sets the units for channel 2 to Volts.
CH2:YUNIT might return CH2:YUNIT "V", indicating that the channel 2 units
are volts.
C commands
74 TBS2000 Series Programmer