11-14
IM DLM6054-01EN
Parameter (Item)
When Searching for Waveform Parameters (Measure)
Selecting the Item That You Want to Search For (Area, Calc, or Select)
• Area1 or Area2: You can select items from the standard waveform parameters. You can select
whether to use the standard waveform parameters in Area1 or Area2.
• Calc:
You can select an item from the expression values registered in the MEASURE
menu.
• Basic:
You can select an item from the standard waveform parameters. You can
select this option when the Mode item in the MEASURE menu is set to Cycle
Statistics.
Search Source Waveform
You can select the search source waveform from the following options.
LOGIC or a channel from CH1 to CH4 or M1 to M4
When Searching for FFT Parameters
Selecting Which Item to Search For (Peak)
• Basic: You can select from the standard FFT analysis items (Peak).
• Calc: You can select from the expressions that have been registered in the FFT menu (which
you can open by pressing SHIFT+MATH/REF).
When Searching for XY Waveform Parameters
You can search through the area values that have been determined using Loop and Polarity
according to the settings in the XY menu (which you can open by pressing SHIFT+DISPLAY).
Upper and Lower Parameter Limits (Upper/Lower)
Settable range for upper and lower limits: –1.0E + 31 to 1.0E + 31
Executing a Search (Exec)
You can search for waveforms that meet the specified search conditions and display only the
waveforms and timestamps that are detected.
Ending the Search (Search Reset)
You can clear the history waveforms that were detected and display all history waveforms.