6-10
IM DLM6054-01EN
6.4 Triggering on an Edge Condition
Procedure
ENHANCED Edge Qualified Menu
Set a trigger using an external signal (EXT) or a channel from CH1 to CH4 as the trigger source.
Press ENHANCED, the Type soft key, the Edge/State soft key, and then the Edge (Qualified) soft
key to display the following menu.
Set the trigger source (CH1-CH4, EXT).
Set the trigger slope ( , ).
Set the trigger level
(when the window comparator is off).
Press to set the qualifications.
When the window comparator is on
Set the window center point.
Set the window width.
Press to configure the trigger coupling, HF rejection,
noise rejection, and the window comparator settings.
Setting the Qualifications (Qualification)
Press the Qualification soft key to display one of the menus shown below. The menu that is displayed
varies depending on the specified trigger source.
When the Trigger Source Is a Channel from CH1 to CH4
You cannot set the trigger source channel.
Set the logic for the qualifications.
Set the qualifications.
Select H, L, or X when the window comparator is off.
Select IN, OUT, or X when the window comparator is on.
When the Trigger Source Is EXT
The same menu appears as that shown above for when the trigger source is from CH1 to CH4.
Because the trigger source is an external signal (EXT), you can specify all of the signal states for
CH1 to CH4 and LOGIC as qualifications.