15-2
IM DLM6054-01EN
15.2 Trigger Output (TRIG OUT)
CAUTION
Do not apply external voltage to the TRIG OUT terminal. Doing so may damage the DL6000/
DLM6000.
External Trigger Output Terminal
When a trigger occurs, the DL6000/DLM6000 produces a TTL level signal. The signal level is normally
high. It becomes low when a trigger occurs.
TRIG OUT
Item Specifications
Connector type BNC
Output level 5 V TTL
Logic Low when a trigger occurs and high after acquisition is completed
Output delay 50 ns or less
Output hold time Low level: 50 ns minimum. High level: 50 ns minimum.
Trigger Output Circuit Diagram and Timing Chart
+5 V
47 Ω
Equivalent to LS06
470 Ω
TRIG OUT
Trigger occurrence (the time at the trigger position)
Trigger output
Internal trigger
Output hold time
Output delay