15-2
IM DLM6054-01EN
15.2  Trigger Output (TRIG OUT)
  CAUTION
Do not apply external voltage to the TRIG OUT terminal. Doing so may damage the DL6000/
DLM6000.
External Trigger Output Terminal
When a trigger occurs, the DL6000/DLM6000 produces a TTL level signal. The signal level is normally 
high. It becomes low when a trigger occurs.
 
TRIG OUT
Item Specifications
Connector type BNC
Output level 5 V TTL
Logic Low when a trigger occurs and high after acquisition is completed
Output delay 50 ns or less
Output hold time Low level: 50 ns minimum. High level: 50 ns minimum.
Trigger Output Circuit Diagram and Timing Chart
 
47 Ω 
Equivalent to LS06 
470 Ω 
TRIG OUT 
 
Trigger occurrence (the time at the trigger position) 
Trigger output 
Internal trigger 
Output delay 
Low Level and High Level Hold Times
 
H
L
Trigger output 
Post-trigger time 
Trigger occurrence 
Trigger occurrence 
Pre
PostPost Pre
Waveform acquisition 
Trigger 
Trigger