17-2
IM DLM6054-01EN
17.2 Messages and Corrective Actions
Messages
Messages may appear on the screen during operation. This section explains the meanings of the error
messages and how to respond to them. You can display the error messages in English or Japanese (see
section 16.1 for details). If servicing is necessary to solve the problem indicated by a message, contact
your nearest YOKOGAWA dealer.
In addition to the following error messages, there are also communication error messages. These
messages are explained in the Communication Interface User’s Manual (IM DLM6054-17EN).
Information
Code Message and Corrective Action Section
51 Turned on pressing the RESET key. All the settings will be initialized. 4.4
52 Firmware is updated. All the settings are initialized.
53 Hardcopy is aborted.
54 File access is aborted.
55 Action-on-trigger is aborted. 6.14
56 Search aborted.
57 Search execution is completed, but no record was found that matched the conditions.
58 Search execution is completed, but no record was found that matched the pattern.
59 Statistical measurement is aborted. Chapter 10
62 The corresponding field was not found.
63 Action-on-trigger is completed. 16.4
64 The instrument is set to remote mode by the communication control.
Press the CLEAR key tochange to local mode.
65 Local lockout is set by the communication control.
To operate using the keys, release the lockout using the communication control.
66 Firmware will be updated. Do you want to proceed?
Note: It will take approx. 5 minutes. Please DO NOT power off the unit until the completion.
Once the procedure is completed, the unit will reboot itself. We recommend you to save the setups
before updating the firmware.
67 Updating Firmware.
Note: Please DO NOT power off the unit. Once the procedure is completed, the unit will reboot itself.
68 Firmware is updated. Will be rebooted.
69 Any serial bus signal can not be detected. Separate
manual*
70 Serial bus automatic setting was aborted. Separate
manual*
71 The symbol/physical value file(.sbl) has not been loaded. 13.9
72 A contradiction in bit numbers of logic setting and symbol definition was detected. Check the symbol/
physical value file(.sbl).
73 Check the input voltage level and attenuation ratio. 5.1
74 Screen will be cleared. Press CLEAR again to proceed. Press ESC to abort. 8.3
* IM DLM6054-51EN
File Errors
Code Message and Corrective Action Section
500 Data size larger than remaining capacity in media.
Delete unnecessary files or use other media.
Chapter 13
501 File does not exist.
Check the file name.
Chapter 13
502 Assigned path does not exist or no media.
Check the path name and media.
Chapter 13
503 Writing prohibited in the media.
Unlock write protection of the media.
Chapter 13