18-18
IM DLM6054-01EN
18.11 General Specifications
Item Specifications
Standard operating conditions Ambient temperature:: 23 ± 5°C
Ambient humidity: 55 ± 10% RH
Supply voltage and
frequency errors:
Within 1% of rating
Warm-up time At least 30 minutes
Storage environment Temperature: –20 to 60°C
Humidity: 20 to 80%RH (no condensation)
Altitude: 3000 m or less
Operating environment Temperature: 5 to 40°C
Humidity: 20 to 80%RH (when the printer is not used; no condensation)
35 to 80%RH (when the printer is used; no condensation)
Altitude: 2000 m or less
Recommended calibration period One year
Rated supply voltage 100 to 120 VAC/220 to 240 VAC(automatic switching)
Permitted supply voltage range 90 to 132 VAC/198 to 264 VAC
Rated supply frequency 50/60 Hz
Permitted supply voltage frequency
range
48 to 63 Hz
Power fuse Built in (not replaceable)
Maximum power consumption 300 VA
Withstand voltage (between the
power supply and case)
1.5 kVAC for 1 minute
Insulation resistance (between the
power supply and case)
External dimensions DL6000
350 mm (W)×200 mm (H)×181 mm (D) when the printer cover is closed; excluding
protrusions
DLM6000
350 mm (W)×200 mm (H)×288 mm (D) when the printer cover is closed; excluding
protrusions
Weight DL6000: Approx. 6.5 kg (Not including options, only the instrument. However, the printer is: Approx. 6.5 kg (Not including options, only the instrument. However, the printer is6.5 kg (Not including options, only the instrument. However, the printer is
included.)
DLM6000: Approx. 7.7 kg ( Not including options, only the instrument. However, the printer: Approx. 7.7 kg ( Not including options, only the instrument. However, the printer7.7 kg ( Not including options, only the instrument. However, the printer
is included.)
Instrument cooling method Forced air cooling; side ventilation
Installation position Horizontal (However, can be installed on a slope using stand and handle,)
Do not install vertically or stack..
Battery backup The clock is backed up with the internal lithium battery.
Battery life: Approx. five years (at an ambient temperature of 25°C)
Environmental protection Uses lead-free soldering
Safety standard Compliant standard
EN61010-1
Overvoltage Category II
1
Measurement Category I
2
Pollution degree 2
3