15-5
IM DLM6054-01EN
Rear Panel Input and Output
15
GO/NO-GO Output Signal
NO-GO OUT Signal
When the determination result is NO-GO, the output signal level (the TTL level) temporarily changes
from high (H) to low (L).
GO OUT Signal
When the determination result is GO, the output signal level (the TTL level) temporarily changes from
high (H) to low (L).
Signal Output Circuit Diagram
GO OUT
NO-GO OUT
100 Ω
4.7 kΩ
100 pF
GO/NO-GO Output Timing
Waveform
acquisition
Waveform
acquisition
Result
Judgment Judgment
Result
NO-GO OUT
GO OUT
EXEC
5 µ or
more
The signal level remains low until the
DL6000/DLM6000 is ready to accept
the next measurement.
If you have specified an action to
perform when conditions are true,
this time is extended until that
action is complete.
15.4 GO/NO-GO Signal Output