2-18
IM DLM6054-01EN
section 6.14 for the procedure
Determination based on specified conditions can be performed at the zone through which waveforms
or automated measurements of waveform parameters pass. If the conditions are met, a given action
can be executed at the same time as the signal acquisition (this does not apply to logic signals). You
can set the action to be carried out to the sounding of an alarm, the saving of measured or screen
capture data, the printing of a screen image, the sending of e-mail, etc.
You can execute the action-on-trigger by pressing the Exec soft key. You cannot use the RUN/STOP
key for this purpose. When action-on-trigger is executed, the trigger mode switches to Normal.
sections 6.15 to 6.17 for the
procedures
The DL6000/DLM6000 uses the result of GO/NO-GO determination to decide whether to perform an
action-on-trigger action. The DL6000/DLM6000 determines whether the acquired waveform meets
the determination conditions (NO-GO) or not (GO). This feature cannot be used on logic signals. The
determination result can be transmitted through the rear-panel GO/NO-GO I/O terminal. The result can
also be used to cause an action-on-trigger action to occur.
This feature is useful for signal testing on electronic device production lines and for tracking down
abnormal phenomena.
You can perform GO/NO-GO determination using:
• A
waveform zone on the screen.
• A rectangular zone on the screen.
• A polygonal zone on the screen.
You can create polygon images on a PC by using the software that comes with the DL6000/
DLM6000.
• A
waveform parameter range.
• A cyclic parameter statistic range.
• An FFT parameter range.
• An XY parameter range.
• A mask test item range.
2.5 Acquisition Conditions