EasyManuals Logo

YOKOGAWA DLM4000 Series User Manual

YOKOGAWA DLM4000 Series
302 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #61 background imageLoading...
Page #61 background image
2-29
IM DLM4038-02EN
FDF Mode (FDF)
Press the Mode soft key and then the FDF soft key to display the following menu.
Set the trigger condition (0(CAN), 1(CAN FD)).
Set the trigger type to FDF.
Setting Trigger Conditions (Condition)
Set the FDF bit state as a trigger condition.
0 (CAN): When the FDF bit is dominant, the DLM4000 assumes that the frame is a CAN bus
signal frame and triggers.
1 (CAN FD): When the FDF bit is recessive, the DLM4000 assumes that the frame is a CAN FD
bus signal frame and triggers.
ESI Mode (ESI (Error Passive))
Press the Mode soft key and then the ESI (Error Passive) soft key.
The DLM4000 triggers when the ESI bit is recessive (error passive).
2.11 Triggering on CAN FD Bus Signals (Option)