CELL-DYN
®
3/35 Sample Loader TS Guide 9140234B-June 1997
TROUBLESHOOTING
1.2 ERROR MESSAGES
1 - 6
1
Search
Book TOC
Go Back
Alarm 4 (s) Sample Tube not sensed correctly. IP - 18
Alarm 4 (t) Safety Cover removed durin
g
sample processin
g
. IP - 19
Alarm 4 (u) Rack Number Misread. Barcode reader did not properly read a
rack barcode.
IP - 20
Autosampler Command Ne
g
atively Acknowled
g
ed Reseat
RS232
Cable At
Both
Ends
And
Restart
Auto Stand-by Mode None
Blood Detected
in Sample Line
If system automatically rinses, and error does not recur. None
If "Blood Detected in Sample Line" error displays on bulletin line. IP - 1
If FATAL error displays on bulletin line. Cycle
Power,
then
IP - 1
Check Rack Alarm None
Cover Open IP - 19
Table 1 - 1 Error Messages(continued)
MESSAGE
CORREC-
TIVE
ACTION