CELL-DYN
®
3/35 Sample Loader TS Guide 9140234B-June 1997
VERIFICATION PROCEDURES
VP - 3.8 Rack to Laser Shield Check
3 - 40
3
Search
Book TOC
Go Back
10.IF Check is OK,
g
o to “VP - 3.10 X-Axis Solenoids to Platen Channel Check” (Pa
g
e 3-43).
11.IF Check Fails,
g
o to “VP - 3.9 Rack to Laser Shield Adjustment” (Pa
g
e 3-41).
Figure 3-17 Rack to Laser Shield Position
--- END ---
SL-165
Rack Guide
Ali
g
nment Tool
Ri
g
ht
Cut-Out
Laser
Shield
#3
#5