CELL-DYN
®
3/35 Sample Loader TS Guide 9140234B-June 1997
VERIFICATION PROCEDURES
VP - 6.4 Mixing Contact Sensor Check/Adjustment
3 - 101
3
Search
Book TOC
Go Back
g
. Verify REPEAT LED Flashers.
CAUTION:Ensure slot in Paddle faces Sensor.
h. Verify Contact Sensor turns Off.
- [CE MARK]
Release the Mixin
g
Paddle. Verify the Sensor is unblocked and REPEAT LED
is Off.
- [NON-CE MARK]
Position Mixin
g
Paddle to block 1/2 of Sensor. Verify REPEAT LED is
Off.
9. If Check Fails:
[CE MARK]
adjust R7 (with Mixin
g
Paddle positioned as indicated in Table 3 - 6, Pa
g
e-3-100).
[NON-CE MARK]
adjust VR6 (DO NOT exceed position shown in Fi
g
ure 3-53, Pa
g
e-3-102).
NOTE:If unable to perform adjustment, add reflective tape or white correction fluid to top of
Mixing Paddle and readjust.
10.Go to “VP - 6.5 Mixin
g
Limit Sensor Check/Adjustment” (Pa
g
e 3-103).