CELL-DYN
®
3/35 Sample Loader TS Guide 9140234B-June 1997
VERIFICATION PROCEDURES
VP - 6.5 Mixing Limit Sensor Check/Adjustment
3 - 105
3
Search
Book TOC
Go Back
E Loader
[NON-CE MARK]
1. Put Sample Loader in Manual Mode.
a. Press and Hold
MAN
and
START
.
b. Turn Power
ON
, hold keys down until BEEPING stops.
2. Place Rack Guide Ali
nment Tool in ri
ht end of platen (Fi
ure 3-55, Pa
e-3-106).
3. Load Mixin
Head Ali
nment Tool in position #4 of rack.
Tool must be fully seated.