3/35 Sample Loader TS Guide 9140234B-June 1997
VERIFICATION PROCEDURES
VP - 6.5 Mixing Limit Sensor Check/Adjustment
3 - 111
3
Search
Book TOC
Go Back
G Loader
1. Put Sample Loader in Manual Mode.
a. Press and Hold
MAN
and
START
.
b. Turn Power
ON
, hold keys down until BEEPING stops.
2. Clear all racks from under mixin
g
paddle.
3. Press (Z
) key down until mixin
g
paddle stops.
4. Plate behind mixin
g
paddle and Sensor Mountin
g
Bracket should clear the laser shield and the
vent wash block by
.020 in. (0.5mm) (Fi
g
ure 3-59).
5. If check fails, adjust Limit Sensor.