3/35 Sample Loader TS Guide 9140234B-June 1997
VERIFICATION PROCEDURES
VP - 13.1 Lead Bubble Check/Adjustment
3 - 155
3
Search
Book TOC
Go Back
d. Aspirate remainin
cleaner by manually openin
Solenoid 96 and release lever from Sample
Peristaltic Pump P2.
Cleaner moves throu
h Shear Valve via Peristaltic Pump to Waste Chamber 2.
NOTE:The typical clot is in the T-fitting that joins the peristaltic pump to the Shear Valve and
Solenoid 92.
e. Run back
round counts to flush cleaner.
5. Load 3 Vacutainer
®
tubes (G Loader - use Sarstedt tube), each with 1-3 ml of control or blood
sample.
6. Enter Normal Mode.
7. Press
START
.