6241A/6242 DC Voltage Current Source/Monitor Operation Manual
1. PREFACE
1-1
1. PREFACE
This manual describes the accessories, operating environment, precautions, and operating check for person-
nel who operate the 6241A/6242. Read this manual before using the 6241A/6242.
1.1 Product Overview
The 6241A/6242 is a DC Voltage Current Source/Monitor with wide source/measurment ranges, as shown
below.
6241A: Voltage 0 to ±32 V, Current 0 to ±500 mA
6242: Voltage 0 to ±6 V, Current 0 to ±5 A
The instruments offer high sensitivity with 4
1
/
2
digit source resolution and 5
1
/
2
digit measurement resolution,
as well as various sweep functions and a pulse measurement function with a minimum pulse width of 50 μs.
They can be widely used as a power source for evaluation and characteristic tests in R&D fields such as semi-
conductors or electrical components.
The 6241A/6242 characteristics are described below.
• Source and Measurement range 6241A: Up to ±32 V, ±500 mA
6242: Up to ±6 V, ±5 A
• Voltage source/measurement range: 6241A: 300 mV to 30 V
6242: 300 mV to 6 V
• Current source/measurement range: 6241A: 30 μA to 500 mA
6242: 30 μA to 5 A
• Source digits/measurement digits: source: 4
1
/
2;
measurement: 5
1
/
2
• Voltage source/voltage measurement resolution: source 10 μV; measurement: 1 μV
• Current source/current measurement resolution: source 1 nA; measurement: 100 pA
• Voltage source current measurement (VSIM) and Current source voltage measurement (ISVM)
• Voltage source voltage measurement (VSVM) and Current source current measurement (ISIM)
• Sink-enabled bipolar output
• Minimum pulse width: 50 μs
• Linear, two-slope linear, fixed, random sweep functions for characteristic test
• Detection functions such as limiter (compliance), overload, and overheat
• Synchronized operation function by combining two or more 6241A/6242 units
• GPIB and USB for integrating an automated measurement system as standard