6241A/6242 DC Voltage Current Source/Monitor Operation Manual
5.2.9 Calculation Functions
5-52
5.2.9.4 Max/Min Calculation
1. Calculation expression
Max/Min calculation determines the maximum, minimum, average, and total values while the calcu-
lation is set to ON.
2. Calculation result
Select the items, H) COMPUTE, 4) View Mx/Mn on the Menu screen to refer to the results.
1. Number of measurement times
2. Maximum value
3. Minimum value
4. Average value
5. Total value
3. Operation
The enabled data except for over range and error data is calculated.
It buzzes when the alarm setting is ON and the maximum value or minimum value is updated.
However, it may buzz even when the displayed data does not change. This is because the meas-
urement resolution is smaller than that of the display.
The calculation is turned OFF by changing the measurement function or executing the *RST
command.
The calculation result is cleared and the calculation is restarted under the following conditions:
1. Switching NULL calculation between ON and OFF
2. Changing NULL data
3. Switching Scaling calculation between ON and OFF
4. Changing Scaling Constant