6241A/6242 DC Voltage Current Source/Monitor Operation Manual
8.1.5 Calibrating Operation
8-11
11. Verify the full-scale calibration value.
XUP and XDN can fine-adjust the calibration coefficient.
12. Proceed to the next step.
When the voltage calibration mode is changed: XNXT
When the operation proceeds to voltage-measurement calibration process: XVM
8.1.5.3 Voltage-measurement Calibration
1. Using XVM, change the mode to voltage measurement calibration mode.
2. Select the range.
300 mV range: XR3
3 V range: XR4
30 V range: XR5
3. Using XDAT, set the DMM data input mode.
4. Set up the DMM read value using XD data.
5. Read out and verify the measurement data.
6. Select XNXT to proceed to the full-scale calibration mode.
7. Set up the DMM read value using XD data.
8. Read out and verify the measurement data.
9. Proceed to the next step.
When the voltage range is changed: XNXT
When the operation proceeds to current-source
and voltage limiter calibration process: XIS, XVLH, XVLL