6241A/6242 DC Voltage Current Source/Monitor Operation Manual
8.1.5 Calibrating Operation
8-12
8.1.5.4 Current-source and Current-limiter Calibration
1. Select the current calibration mode.
Current-source: XIS
Current HI limiter: XILH
Current LO limiter : XILL
2. Select the range.
30 μA range: XR-1
300 μA range: XR0
3 mA range: XR1
30 mA range: XR2
300 mA range: XR3
500 mA range: XR4
3. Perform the same procedure as described in Section 8.1.5.2, “Voltage-source and
Voltage-limiter Calibration.”
4. Proceed to the next step.
When the current calibration mode is changed: XNXT
When the operation proceeds to current-measurement
calibration process: XIM
8.1.5.5 Current-measurement Calibration
1. Using XIM, change the mode to current measurement calibration mode.
2. Select the range.
30 μA range: XR-1
300 μA range: XR0
3 mA range: XR1
30 mA range: XR2
300 mA range: XR3
500 mA range: XR4
3. Perform the same procedure as described in Section 8.1.5.3,
“Voltage-measurement Calibration.”