6241A/6242 DC Voltage Current Source/Monitor Operation Manual
8.2.5 Calibrating Operation
8-26
11. Verify the full-scale calibration value.
XUP and XDN can fine-adjust the calibration coefficient.
12. Proceed to the next step.
When the voltage calibration mode is changed: XNXT
When the operation proceeds to voltage-measurement
calibration process: XVM
8.2.5.3 Voltage-measurement Calibration
1. Using XVM, change the mode to voltage measurement calibration mode.
2. Select the range.
300 mV range : XR3
3 V range: XR4
6 V range: XR5
3. Using XDAT, set the DMM data input mode.
4. Set up the DMM read value using XD data.
5. Read out and verify the measurement data.
6. Select XNXT to proceed to the full-scale calibration mode.
7. Set up the DMM read value using XD data.
8. Read out and verify the measurement data.
9. Proceed to the next step.
When the voltage range is changed: XNXT
When the operation proceeds to current-source and
voltage limiter calibration process: XIS, XVLH, XVLL
8.2.5.4 Current-source and Current-limiter Calibration (30 µA to 300 mA)
1. Select Current Calibration mode.
Current-source: XIS
Current HI limiter: XILH
Current LO limiter: XILL
2. Select the range.
30 μA range: XR-1
300 μA range: XR0
3 mA range: XR1
30 mA range: XR2
300 mA range: XR3
3. Perform the same procedure as described in Section 8.2.5.2, “Voltage-source and