6241A/6242 DC Voltage Current Source/Monitor Operation Manual
2.2.4 Initializing Setting Conditions
2-34
2.2.4 Initializing Setting Conditions
The following procedure returns the 6241A/6242 to the factory settings.
However, the following items cannot be initialized.
• Selected interface
• GPIB address
• Talk only/Addressable
• Header output
•USB.Id
• Load Parameters at power on
Operation Character display area
1.
Press MENU and press or to select K)
PARAMETER.
K) PARAMETER
2.
Press to go to the Select level.
1) Parm Load
Load0
3.
Press to go to the Input/Run level.
1) Parm Load
Ent Load0
4.
Rotate to select “Ld Dflt”.
1) Parm Load
Ent Ld Dflt
5. Press ENTER.
6. When loading is complete, “Done” is displayed and
the menu reverts to the Select level.
1) Parm Load
Done
7. Press EXIT to exit the menu. 1) Parm Load
Ld Dflt