6241A/6242 DC Voltage Current Source/Monitor Operation Manual
5.2.14 Error log
5-65
5.2.14 Error log
The 6241A/6242 holds error numbers in the error log memory when it detects an error.
1. Operation
A maximum of 5 memory areas are available for the error log and they operate as follows:
2. Clearing an error log
The error logs are cleared by the following and the ERR indicator is extinguished.
•Power ON
• When reading error log (This error log is cleared after displaying the error log screen and com-
pleting the Menu screen.)
• When the ERL? or *CLS command is executed.
Not cleared by *RST.
3. Reading error log
Select the items, M) SYSTEM, 5) Error Log on the Menu screen.
The number of errors is displayed.
Press to display the error content display screen.
Turn to change the error log number.
An error message such as the one below appears if the entered value is set out of range.
• A maximum 5 error numbers are stored in the order of detection.
• If detection exceeds more than 5, then the fifth error log is overwritten by the last error.
• The ERR indicator illuminates when an error log is stored.
5
4
3
2
1
Order of
storing
Overwrites the 6th and
later error here.
Number of error logs
Error log