6241A/6242 DC Voltage Current Source/Monitor Operation Manual
2.3.3 Loading Parameters
2-49
2.3.3 Loading Parameters
Operation Character display area
1.
Press MENU and press or to select K)
PARAMETER.
K) PARAMETER
2.
Press to go to the Select level.
1) Parm Load
Load0
3.
Press or to select 1) Parm Load.
1) Parm Load
Load0
4.
Press to go to the Input/Run level.
1) Parm Load
Ent Load0
5.
Rotate to select the load area.
1) Parm Load
Ent Load2
6. Press ENTER.
7. When loading is complete, “Done” is displayed and
the menu reverts to the Select level.
1) Parm Load
Done
8. Press EXIT to exit the menu. 1) Parm Load
Load2