6241A/6242 DC Voltage Current Source/Monitor Operation Manual
1.4 Operating Environment
1-4
1.4 Operating Environment
This section describes the required environmental and power supply conditions.
1.4.1 Environmental Conditions
The 6241A/6242 must be installed in an environment meeting the following conditions:
Ambient temperature: 0ºC to +50ºC (temperature range for operation)
-25ºC to +70ºC (temperature range for storage)
Relative humidity: 85% or lower (without condensation)
Location not subject to corrosive gasses
Away from direct sunlight
Dust free
Vibration free
Noise free
The 6241A/6242 is designed with full consideration given to the noise contained in the AC power
line. Nevertheless, it is recommended that the 6241A/6242 be used in an environment with as little
line noise as possible.
If line noise is unavoidable, use a noise filter.
Positioning of the 6241A/6242
A cooling fan is located on the rear panel and vents are located on the side panels. Do not block the
fan and vents. Leave at least 10 cm of free space between the rear panel and the wall. Also, do not
position the 6241A/6242 in a position with the rear panel facing down.
Obstructing the vents will cause the internal temperature to rise, possibly causing faulty operation.
Mounting in a rack
Ensure that exhaust air from other devices is not directed at the vents on the side of the 6241A/6242.
To prevent the temperature in the rack from rising, install a heat sink fan.