4.2.15 VS/IS Key (Source Function)
4-18
4.2.15 VS/IS Key (Source Function)
Selects source functions, voltage source or current source. The unit indicates VS or IS mode. When chang-
ing between VS and IS, the following operations are restricted:
• Cannot be changed during sweep operation when the source mode is Sweep.
• Forces into Suspend during sweep stop (start trigger wait status) when the source mode is Sweep.
• Forces into Suspend when executing during an operation in the DC source/pulse source mode.
4.2.16 123... Key (Direct Input Mode)
This key is used to switch into a direct input mode that can input numeral data and display the setting value
at half-brightness.
In this status, green colored functions on the panel are enabled.
However, EXP key is only enabled when Disp Unit is the exponent and a calculation is executed. Press
ENTER to apply the input numerical data and to release the direct input mode.
Press EXIT to revert to the value when shifted to the direct input mode and release the direct input mode.
4.2.17 4W/2W Key (Remote Sensing Selection)
This key selects output sensing 4-wire or 2-wire connection.
4W: Sets output sensing to 4-wire connection. The 4W indicator turns on.
2W: Sets output sensing to 2-wire connection. The 2W indicator turns on.