6241A/6242 DC Voltage Current Source/Monitor Operation Manual
9.2 Source and Measurement Function
9-16
9.2 Source and Measurement Function
DC source and measurement: Source/measurement of DC voltage/current
Pulse source and measurement: Source/measurement of pulse voltage/current
(However, measurement auto range in Pulse source is impossible.)
DC sweep source and measurement: Source and the measurement with Linear, Two slope Linear, Random
and fixed level.
Pulse Sweep source and measurement:
Source and the measurement with Linear, Two slope Linear, Random
and fixed level. (However, measurement auto range in Pulse source is
impossible.)
Integration time: Select from 100 μs; 500 μs; 1 ms; 5 ms; 10 ms; 1 PLC; 100 ms; 200
ms, or S/H.
S/H: Measuring in Sample Hold Mode (Integration time: 100 μs)
(Enabled only in the pulse source and pulse sweep source
modes.)
(PLC: Power Line Cycle, 50 Hz: 20 ms, 60 Hz: 16.66 ms)
Sweep mode: Reverse ON (forward to backward)/OFF (one way)
Number of times of repeating Sweep:1 to 1000 times or infinite
Maximum number of step for Sweep:8000 steps
Random sweep maximum memory: 8000 data items
The measurement data memory: 8000 data items
Measurement auto range: Available only in VSIM or ISVM
Measurement function link mode: Links the source function to the measurement function. (VSIM or
ISVM) ON/OFF available
Limiter: It is possible to set separately at HI and LO sides.
(However, in case of current-limiter, the same polarity can not be set.)
Calculation function: NULL calculation
comparator calculation (HI, GO, or LO)
Scaling calculation
MAX, MIN, AVE, TOTAL calculation
Trigger style: Auto-trigger, External-trigger
Output terminal: Front; Safety socket
HI OUTPUT, HI SENSE, LO OUTPUT, and LO SENSE
The maximum voltage applied between the terminals:
6241A: 32 V peak Max (between HI-LO)
6242: 6 V peak Max (between HI-LO)
2 V peak Max (between OUTPUT and SENSE)
250 V Max (between LO and chassis)
Maximum remote sensing voltage: ±1 V Max; Between HI OUTPUT - HI SENSE and LO OUTPUT -
LO SENSE
(the output voltage between HI SENSE and LO SENSE
must be within the maximum output voltage range)