6241A/6242 DC Voltage Current Source/Monitor Operation Manual
6.3.3 Setting GPIB
6-7
6.3.3 Setting GPIB
These settings are enabled if the GPIB interface is selected.
• Setting the Address
• Talk only setting
Operation Character Display Area
1.
Press MENU and press or to select
L) I/F.
L) I/F
2.
Press to go to the Select level.
1) I/F BUS
GPIB
3.
Press or to select 2) GPIB.
(Current address setting)
2) GPIB Adr
01
4.
Press to go to the Select level.
2) GPIB Adr
01
5.
To set the address, press or to select the
digit to change and use to increase or decrease
the value. Alternatively, directly enter the address
value pressing the 123... key.
2) GPIB Adr
17
6. Press EXIT to exit the menu.
Operation Character Display Area
1.
Press MENU and or to select L) I/F.
L) I/F
2.
Press to go to the Select level.
1) I/F BUS
GPIB
3.
Press or to select 4) Talk Only.
(Current address)
4) Talk Only
Off
4.
Press to go to the Select level.
4) Talk Only
Off
5.
Rotate to set the Talk Only ON or OFF.
4) Talk Only
ON
6. Press EXIT to exit the menu.