6241A/6242 DC Voltage Current Source/Monitor Operation Manual
6.7.4 TER? Command
6-40
6.7.4 TER? Command
The TER? command reads out the self test result.
1. Command response
2. Description of values a, b, c, d, and e
The TER register column in Table 5-17 shows the error causes and register values of a, b, c, d and e.
If a VSVM3 V + FS error occurs during the self test, the command response is as follows.
00000, 00000, 00016, 00000, 00000
Source
Operating or
Standby
H Output is set to OFF (Standby).
E Output is set to ON (Operating).
E?, H? Responds to the current output status.
Response:
Random sweep
memory data
(using D com-
mand)
N [adr] The random-sweep memory-data starts from N
command and completes at P command.
0
*6
P N<adr>,D<data1><UNIT>,D<data2>
<UNIT>,...,P
adr:Memory address (0
to 7999)
data1:adr address voltage or cur-
rent-source value
data2:adr + 1 address voltage or cur-
rent-source value
*6: It is not initialized by RINI or *RST command.
Item Command Description
Default Operation
Power
ON
Default
setting
During DC/
pulse
operation
During
sweep
operation
Status E?, H?
Operating E
Suspended H
Standby H
Note:
1. Not specifying the source range causes the FIT
range to be specified.
2. A source value different from the current-source
function cannot be set.
ddddd,ddddd,ddddd,ddddd,ddddd
a
bc
d
e