6241A/6242 DC Voltage Current Source/Monitor Operation Manual
5.2.2 Pulse Source Mode Operation
5-10
5.2.2 Pulse Source Mode Operation
Table 5-3 below shows pulse source mode operation.
Table 5-3 Pulse Source Mode Operation (1/2)
Operational
condition
Trigger mode Explanation Operation Remarks
Operate
ON
AUTO
Executes consec-
utive measure-
ment with the set
period time Tp.
Tp: Period Time
Td: Measurement
Delay Time
Tds:Source Delay Time
Tm: Measurement Time
(Integration time +
Measurement data
processing time)
Tcn: Operation Processing Time
Trc: Range Change Processing
Time
HOLD
Executes mea-
surement after
trigger input.
Source value
changed
AUTO
Changing source
value does not
induce changing
range.
HOLD
Standby status
Operate
Tp Tp
Tw
Tw
Td
Tm
Td
Tm
Tcn
Tds Tds
COMPLETE OUT [FRONT]
COMPLETE OUT [END]
COMPLETE OUT [HI/GO/LO]
SYNC OUT
Standby status
Tp
COMPLETE OUT [FRONT]
TRIGGER IN
Operate
Tcn
Tw
Td
Tm
Tds
Measurement
trigger
SYNC OUT
COMPLETE OUT [HI/GO/LO]
COMPLETE OUT [END]
Tp Tp
Tw
Tw
Td
Tm Tm
Td
Tds Tds
Base value
Source value change
proc essing time
(Range not change)
COMPLETE OUT [FRONT]
SYNC OUT
COMPLETE OUT [HI/GO/LO]
COMPLETE OUT [END]
Source value change
processing time
(Range not change)
Td
COMPLETE OUT [FRONT]
COMPLETE OUT [HI/GO/LO]
COMPLETE OUT [END]
Tp
Tw
Tm
Tds
Base value
Source value changed
(Range not change)
SYNC OUT
TRIGGER IN
Measurement
trigger

Table of Contents