S:\Hp8960\E1962B CDMA 2000\Pi Release\Reference Guide\Chapters\cdma2000_prog_afan_freq.fm
Programming an Audio Frequency Measurement
Programming an Audio Frequency Measurement
This section provides an example of how to make an audio frequency measurement with the Audio Analyzer
(AFANalyzer) via GPIB.
The following procedure assumes that an audio source is connected to the AUDIO IN connectors. See Audio
Analyzer Measurement Description on page 38.
1. Configure audio analyzer measurement parameters using the SETup subsystem. When more than one
audio frequency signal may be present, use the 100 Hz bandpass filter to isolate the desired signal for
measurement (see the Programming Example below).
2. Start the audio analyzer measurement using the INITiate subsystem.
3. Use the FETCh? subsystem to obtain audio analyzer measurement results.
Programming Example
10 OUTPUT 714;SETup:AFANalyzer:CONTinuous OFF !Set the audio analyzer
20 !measurements to single trigger mode.
30 ! OUTPUT 714;SETup:AFANalyzer:PEAK:VOLTage 3V !Set the Expected peak voltage
40 !for the audio input level in Vpeak.
50 OUTPUT 714;SETup:AFANalyzer:FREQuency:STATe ON !Enable audio frequency meausurement.
60 OUTPUT 714;SETup:AFANalyzer:FILTer TBPass !Select the 100 Hz bandpass filter.
70 OUTPUT 714;SETup:AFANalyzer:FILTer:TBPass 6000 Hz !Set the 100 Hz bandpass
80 !filters center frequency to 6 kHz (example; to measure SAT frequency).
90 OUTPUT 714;SETup:AFANalyzer:TIMeout 3S !Set a timeout value of 3 seconds
100 !in case the measurement cannot be made.
110 OUTPUT 714;INITiate:AFANalyzer !Start the AF Analyzer measurement.
120 OUTPUT 714;FETCh:AFANalyzer:FREQuency? !Fetch the Audio Frequency result.
130 ENTER 714;Audio_frequency !Enter the returned value into a variable.
140 END
Related Topics
Audio Frequency Measurement Description on page 42
AFANalyzer Troubleshooting on page 704
Audio Analyzer Measurement Description on page 38
INITiate on page 476
SETup:AFANalyzer on page 505
FETCh:AFANalyzer on page 415

Table of Contents

Other manuals for Agilent Technologies 8960 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8960 Series and is the answer not in the manual?

Agilent Technologies 8960 Series Specifications

General IconGeneral
BrandAgilent Technologies
Model8960 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals