Gated Power Measurement Troubleshooting
Gated Power Measurement Troubleshooting
Possible Setup Issues
IMPORTANT If the Access Probe Power measurement is currently selected and the Trigger Arm setting is
Continuous, all other measurements are disabled. Use the SETup:CAPPower:CONTinuous
command to set the trigger arm state to single and retry this measurement.
During remote operation of this measurement, you should configure the Trigger Arm to Single. See
SETup:GPOWer:CONTinuous . If single triggering is used during manual operation, press the START
SINGLE key to trigger each measurement.
The test set receivers expected signal amplitude and channel number (or frequency) must be set up correctly
for accurate measurements. This is normally done automatically (see Active Cell Operating Mode ) but can
also be done manually (see Receiver Control on page 194) .
The following settings are required for the gated power measurement:
Operating Mode must be Active Cell
System Type must be IS-95 or IS-2000 with Radio Config set to (Fwd1, Rvs1) or (Fwd2, Rvs2)
FCH Service Option Setup must be set to SO2 or SO9 (loopback service option)
Traffic Data Rate must be set to Eighth.
The input signal must conform to Measurement Requirements .
Related Topics
Integrity Indicator on page 222
Gated Power Measurement Description

Table of Contents

Other manuals for Agilent Technologies 8960 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8960 Series and is the answer not in the manual?

Agilent Technologies 8960 Series Specifications

General IconGeneral
BrandAgilent Technologies
Model8960 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals