EasyManuals Logo

Agilent Technologies 8960 Series Reference Guide

Agilent Technologies 8960 Series
786 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #63 background imageLoading...
Page #63 background image
63
S:\Hp8960\E1962B CDMA 2000\Pi Release\Reference Guide\Chapters\cdma2000_meas_hwq_desc.fm
Handoff Waveform Quality Measurement Description
Measurement Requirements
Setup:
The following settings are required for the handoff waveform quality measurement:
Operating Mode must be Active Cell
System Type must be IS-2000 with Radio Config set to (Fwd3, Rvs3) or (Fwd4, Rvs3)
A call must be connected
Input Signal:
The handoff waveform quality measurement meets or exceeds specifications when the following input signal
requirements are met:
The frequency of the signal being measured must be in the range of 412 MHz to 483 MHz, 800 MHz to 960
MHz, or 1.7 GHz to 2.0 GHz, and within 100 kHz of the expected frequency.
The signal level into the test sets RF IN/OUT connector must be in the range of -25 dBm to +37 dBm, and
within 9 dBm of the expected input power (see RFANalyzer:AUTO:POWer[:SELected]? ) to meet
specifications.
Related Topics
Programming a Handoff Waveform Quality Measurement
SETup:HWQuality
FETCh:HWQuality
INITiate
READ
ABORt
Test Adherence to Standards
Handoff Waveform Quality Measurement Troubleshooting

Table of Contents

Other manuals for Agilent Technologies 8960 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8960 Series and is the answer not in the manual?

Agilent Technologies 8960 Series Specifications

General IconGeneral
BrandAgilent Technologies
Model8960 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals