S:\Hp8960\E1962B CDMA 2000\Pi Release\Reference Guide\Chapters\cdma2000_hpib_system_error.fm
SYSTem:ERRor?
SYSTem:ERRor?
Diagram Conventions on page 248
SYSTem:ERRor?
Function Queries the contents of the error/event queue. The error/event queue may contain one or more
messages with an error or event description.
Manual users may view the message log from the SYSTEM CONFIG screen. The contents of the
error/event queue and the message log may not match. For example, manual user errors are not
displayed with SYSTem:ERRor? They are viewed from the Message Log. See Error Messages .
Query Error/Event Queue
Range: 0 to 100 messages up to 255 characters in length
Programming Example
OUTPUT 714;"SYSTEM:ERROR?" !Query the contents of the Error/Event Queue
SYSTem :ERRor?
(returns num value, string)