S:\Hp8960\E1962B CDMA 2000\Pi Release\Reference Guide\Chapters\cdma2000_hpib_call_rfgen.fm
CALL[:CELL]:RFGenerator
CALL[:CELL]:RFGenerator
Diagram Conventions on page 248
CALL[:CELL]:RFGenerator:FREQuency
Function Sets/queries the RF generator CDMA forward channel frequency. If no units are specified the
setting defaults to Hz.
The test set will tune to the frequency entered using this command when the
CALL[:CELL]:CONTrol:DOWNlink:FREQuency:AUTO command is set to manual.
Setting Range: 292 MHz to 2700 MHz
Resolution: 0.1 Hz
Query Range: 292 MHz to 2700 MHz
Resolution: 0.1 Hz
*RST Setting 851 MHz
Programming Example
OUTPUT 714;"CALL:CELL:RFGENERATOR:FREQUENCY 450 MHZ"
!Sets the CDMA forward channel frequency to 450 MHz.
CALL
:RFGenerator
? (returns num value)
[:CELL]
:FREQuency
<sp><num value>[HZ|KHZ|MHZ|GHZ]