595
S:\Hp8960\E1962B CDMA 2000\Pi Release\Reference Guide\Chapters\cdma2000_hpib_status_preset.fm
STATus:PRESet
STATus:PRESet
Diagram Conventions on page 248
STATus:PRESet
Function Presets the status subsystem.
Presetting the status subsystem performs the following functions:
All enable registers are set to 0 (not enabled)
All Positive Transition Registers (PTRs) are set to 1 (positive transitions enabled).
All Negative Transition Registers (NTRs) are set to 0 (negative transitions disabled).
XXXXXXXXX XXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXX
STATus :PRESet