554
S:\Hp8960\E1962B CDMA 2000\Pi Release\Reference Guide\Chapters\cdma2000_hpib_setup_hwquality.fm
SETup:HWQuality
SETup:HWQuality
Diagram Conventions on page 248
SETup:HWQuality:TIMeout[:STIMe]
SETup:HWQuality:TIMeout:STATe
Function Sets the handoff waveform quality measurements timeout value, and sets the waveform quality
measurements timeout state to 1 (on).
The query form returns the waveform quality measurements timeout value.
Setting Range: 1 to 999 seconds
Resolution: 1 second
Query Range: 1 to 999
*RST Setting 10 seconds, on
Programming Example
OUTPUT 714;"SETUP:HWQUALITY:TIMEOUT:STIME 5 S"
!Sets timeout state to ON and sets the timeout time to 5 seconds.
Function Sets/queries the handoff waveform quality measurements timeout state.
Setting Range: 1|ON|0|OFF
Query Range: 1|0
*RST Setting 0 (off)
Programming Example
OUTPUT 714;"SETUP:HWQUALITY:TIMEOUT:STATE ON"
SETup :HWQuality
:TIMeout
[:STIMe]
? (returns num value)
:TIME
? (returns num value)
:STATe
<sp>1|ON|0|OFF
Complex Command
? (returns 1|0)
<sp><num value>[S|MS|US|NS]
<sp><num value>[S|MS|US|NS]

Table of Contents

Other manuals for Agilent Technologies 8960 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8960 Series and is the answer not in the manual?

Agilent Technologies 8960 Series Specifications

General IconGeneral
BrandAgilent Technologies
Model8960 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals