Page #141 background imageLoading...
Page #141 background image
141
S:\Hp8960\E1962B CDMA 2000\Pi Release\Reference Guide\Chapters\cdma2000_prog_generic_step6_ac.fm
Step 6: Reconfigure Test Set and Mobile Station Connection Parameters
Step 6: Reconfigure Test Set and Mobile Station Connection Parameters
The following information provides additional details on Step 6 of the Programming Flowchart. This
information is applicable to all test applications.
Description
After performing a set of measurements on the mobile station using the configuration established in step 2,
you may want to change this configuration and test the mobile station again. This step involves changing
testing conditions such as channel, mobile station transmit power level, or cell power.
Contents
Change Channels
Change Other Connection Parameters
Change Channels
In a manufacturing environment it is common to test a mobile station on more than one channel. Some or all
of the tests may be performed on multiple channels, to verify the mobile stations performance in a range of
operating frequencies.
Perform a Handoff (Handover) in Active Cell Mode
If you are operating in active cell mode, to change channels you perform a handoff (handover) to the new
channel. The control program sends the commands to initiate the handoff (handover), and then determines
whether the transition was successfully completed. This is accomplished using the CALL:STATus? query.
The CALL:STATus? query immediately returns the state of the call at the time the query is received.
After issuing the commands to perform a handoff (handover), you should send the CALL:STAT? query to verify
that the call is still connected before resuming testing.
Example Responses Returned from the CALL:STAT? Query
Response Meaning
IDLE The call is in the Idle state
ALER The call is in the Alerting transitory state.
HAND The call is in the Handoff transitory state.
PAG The call is in the Paging transitory state.
CONN The call is in the Connected state.

Table of Contents

Other manuals for Agilent Technologies 8960 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8960 Series and is the answer not in the manual?

Agilent Technologies 8960 Series Specifications

General IconGeneral
BrandAgilent Technologies
Model8960 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals